甲醛检测管NO.91P NO.91P

产品价格: < 面议
产品型号:NO.91P
 牌:日本GASTEC
公司名称:气体检测仪器网|北京圣达骏业科技有限公司
  地:北京丰台
浏览次数:
更多
立即询价 在线咨询

产品简介

甲醛检测管NO.91P,GASTEC甲醛检测管NO. 91P电泵用气体检测管检测范围0.02-1.44ppm

产品详细信息

甲醛检测管NO.91P

甲醛检测管NO.91P

简单介绍
甲醛检测管NO. 91P,GASTEC甲醛检测管NO. 91P气体检测管,甲醛HCHO检测管NO. 91P,GASTEC甲醛检测管NO. 91P电泵用气体检测管检测范围0.02-1.44ppm。
甲醛检测管NO. 91P的详细介绍
甲醛检测管NO. 91P
甲醛检测管NO. 91P,GASTEC甲醛检测管NO. 91P气体检测管,甲醛HCHO检测管NO. 91P,GASTEC甲醛检测管NO. 91P电泵用气体检测管检测范围0.02-1.44ppm。
GASTEC为了满足用户在气体检测方面的需要,而开发出了电子自动采气泵。用传统的电子吸气泵几乎不能进行有效的采气,因为检测管里的填充试剂密度很高。而GASTEC开发的GSP电泵系列能提供强劲吸力,能满足特定的检测管的取样要求。电泵和检测管组成的新的检测系统能适用于作业环境的检测。

Formaldehyde

NO.91P

HCHO  
Specifications
Calibration Range 0.02-0.4 ppm
Measuring range 0.02-1.44 ppm
Sampling (flow) Rate 200mL/min
Sampling Volume 6000mL, 2000mL
Sampling Time 30min, 10min.
Detecting Limit 0.01 ppm
Colour Change Yellow → pink
Corrections for temperature/humidity Temperature
Shelf life 1 year (in the refrigerator)
Accuracy CV=5-10%
(CV: Coefficient of variation = σ: standard deviation / mean value x 100)
Possibly coexisting substances and their interferences
Substance(s) Interference Colour change by itself
Ammonia 2 ppm (or less) No No
Nitrogen dioxide 1 ppm (or less) No No
Acetaldehyde   "+" error
(higher reading)
produces pinkish discolouration
Acetone 0.3 ppm (or less) No Produces pinkish stain around tube opening (inlet)
Ethanol   No No
Ethyl acetate   No No
p-dichlorobenzene   No No
Toluene   No No

Scrubber 1 (black): removes acetone

Scrubber 2 (purple):
Ammonia and nitrogen dioxide are removed. The presence of ammonia causes the purple scrubber 2 to turn yellow.

The presence of nitrogen dioxide causes scrubber 2 to turn a darker shade of purple
Calibration gas generation Diffusion tube method (analysis: by absorptiometry, acetylacetone method)
Miscellaneous
Electric sampling pump
This detector tube is suited for measuring formaldehyde in indoor environments. The GSP-200 / GSP-300FT series automatic gas samplers (or comparable sampling devices/pumps) should always be used for optimal performance.
A deviation of +/-10% in the sampling flowrate (i.e. amount of air flowing through the tube at a given moment) or the sampling amount (total sampled amount of target substance) causes an average margin of error of approx. +/-10% (maximum +/-20%) in the indicated value.
Formaldehyde indoor concentration threshold: 0.08 ppm (mean value during 30 minutes)
TLV-TWA TLV-STEL Explosive range
-- -- C0.3ppm 7.0×73%

在线询价

您好,欢迎询价!我们将会尽快与您联系,谢谢!
  • *产品名称:
  • 采购数量:
  • 询价有效期:
  • *详细说明:
  •   100
  • *联系人:
  • *联系电话:
  • 附件: 附件支持RAR,JPG,PNG格式,大小在2M范围
验证码: 点击换一张
 
   温馨提示:为规避购买风险,建议您在购买前务必确认供应商资质与产品质量。

   免责申明:以上内容为注册会员自行发布,若信息的真实性、合法性存在争议,平台将会监督协助处理,欢迎举报
产品标签