超声波硬度计 SonoDur3
产品简介
NewSonic SonoDur3超声波硬度计是升级版,NewSonic SonoDur3超声波硬度计坚固耐用 易于使用 快速而准确 德国制造。
产品详细信息
The New UCI Generation after SonoDur2
Mobile Hardness testing onmetals with UCI * in productionand maintenance. Starting withgoods income inspection untilthe fi nal product is reached.
*Ultrasonic Contact Impedance
Norms: ASTM A 1038 and DIN 50159-1, 2
Conversions: ASTM E 140-13, EN ISO 18265
Multiple Solutions for theSonoDur Family and also forMIC10/MIC20 (by probe retrofi t)
NewSonic SonoDur3超声波硬度计
Motor probes (HV 0,1 – HV 0,8): 0.1 kgf, 0.3 kgf, 0.8 kgf
Handheld probes (HV 1 – HV 10): 1 kgf, 3kgf, 5 kgf, 10 kgf
The better Way to Hardness Testing
Rugged, easy to use, fast and precise. Made in Germany.
Instrument Features
• Carrying strap, edge protection, IP65, MIL-810G (vibration, shock, drop)
• Brilliant color display in 5’’, smooth glass shield for easy cleaning
• Direct function access just by touch, intuitive menu control, ideal forMIC10/MIC20 users and others
• No set-up needed at beginning of shift – ready for immediate use
NewSonic SonoDur3超声波硬度计
Android Operating System and Data Transfer
• USB: transfer of measurement data in txt-format using thefile explorer
• WLAN: automatic transfer of single values, completed data stets
• Bluetooth: manual transfer of a set of measurement data
•“Unlimited”storage capacity for measurement data and settings
Intelligent Management ofMeasurement Data and Settings
• Simple access to data and qualified back-traceableassessment of results
• File names can be predefined, automatic closingof data sets
•“AllMeas”- summarized list of results in storeddata sets
Reliable MeasurementTechnique with long-Term Stability
Efficient working tool, optimized for the
daily needs of the inspector
Unique
Can be calibrated in full according to
DAKKS after DIN 50159-2 (calibration
curve 150 – through 900 HV, diamond,
nominal test force)
NewSonic SonoDur3超声波硬度计
Measuring Specifications
Measuring principle |
UCI Method, corresponds to DIN 50159, ASTM A1038 |
||
Test indenter |
Vickers diamond 136° |
||
Test loads Newton scale (1kgf = 9.81 N) |
Motor probes: 1N (0.1 kgf), 3N (0.3kgf) and 8.6 N (0.9 kgf) Handheld Probes: 10N (1 kgf), 30N (3kgf), 49N (5kgf), 98N(10kgf), (Other test loads on request) |
||
Hardness scales and range (according to relevant standards), in this casetable A1 respectively T1, T2 (low alloy steel). Different measuring ranges are valid for othermaterials. When exceeding the limits theconversion range will be extended. The calcu-lated values are highlighted in red besides theoriginal data in HV.
Note: Conversions are acc. to latest ASTM E140-12bԐ1(2013) und EN ISO 18265:2014. Conversions into tensile strength: 98N (10kgf)test load only. |
Vickers Brinell Rockwell Rockwell Rockwell Rockwell Rockwell 5,6 Rockwell (EN ISO 18265 only) Rockwell Knoop (ASTM E140 only) Shore (ASTM E140 only) Tensile strength |
HV HB HRB HRC HRE HRF HRA HRD HR45N HK HS MPa |
10 – 1999 (9999) 76 – 618 41 – 105 20,3 – 68 70 – 108,5 82,6 – 115,1 60,7 – 8 40,3 – 76,9 19,9 – 75,4 87 – 920 34,2 – 97,3 255 – 2180 |
Measurement uncertainty* |
< 4 % (HV5, HV 10). For other test loads and ranges see table below. |
||
Relative repeatability* |
< 5 % (HV5, HV 10). For other test loads and ranges see table below. |
* exceeds DIN 50159, dependent on test load and range (see table below). Specifications are valid for 5 measurements usingVickers reference blocks and according to test conditions given in standard DIN 50159.
NewSonic SonoDur3超声波硬度计
Scope of delivery
SonoDur3, Hardness Tester with Data Logger and Data Export, Data Transfer to PC (USB, WLANor Bluetooth resp.), incl. SONO3-NG, Power Source, SONO2-NG/USB USB-Cable, SONO2-HM, ca. 1,5 mProbe Connection Cable, SONO3-TK-1, Transportation Case, SONO-CD, Product-USB-Stick, SONO3Protect, Protection Foils
Mandatory accessory
Motor probe or handheld probe respectively
Other accessories
Reference blocks (MPA, company certificate),test stands, guiding harnesses, auxiliary SW.
Example for configurations
温馨提示:深圳泰立仪器代理销售德国NewSonic超声波硬度计,因美国GE 硬度计MIC10已停产,
好消息:德国NewSoni新一代超声波硬度计SonoDur3设备完全替代美国GE硬度计并兼容已停产老款探头.
提供可连接 MIC10 和 MIC10DL 的硬度计
SONO-201 负载力 10N(1kg)
SONO-205 负 载力 49N(5kg)
SONO-2010 负载力 98N(10kg)
注意选用此系列探头搭配MIC10及MIC10DL需要更换以下探头线SONO-N1050才可以正常使用:
MIC10探头线
SONO-M1 (0.5米 GE电动探头配合支架适用)
SONO-M10M (1.5米 GE电动探头)
SONO-M1C1050 1.5米 ( GE手动探头适用)