e-201-c-9PH复合电极英文版说明书

分享到:
点击量: 199270 来源: 上海越磁电子科技有限公司

Model E-201-9 Combination Electrode Operation Instruction

E-201-9型复合电极使用说明

I. Application用途

Model E-201-9 electrode is combined by glass electrode and reference electrode for measuring Ph value in water. It is widely used in chemical light industry, pharmaceutical industry, dye-works and scientific institute where pH value needs measuring.

本电极是玻璃电极和参比电极组合在一起的复合电极,用于测量水溶液的pH值。它广泛应用于化轻工业、制药工业、染色工业和科研机构中需要检测水溶液PH值的场合。

II. Specification 规格

1. Measuring range测量范围: 0-14 pH

2. Measuring temperature测量温度: 0-60

3. Zero potential零电位: 7±0.5pH(25)

4. Slope斜率: (PTS)≥98.5%(25)

5. Internal resistance内电阻:≤250MΩ(25)

6. Alkali error碱误差 : 0.2pH(1M Na+  pH14) (25)

7. Response time响应时间: towards 95% equilibrium value, less than 1 sec./对于 95%均衡值,少于1秒。

III. Operation and maintenance使用和维护

1. Before measuring, electrode must be calibrated with standard buffer solution.

测量前,电极必须用标准缓冲液校准。

2. Sensitive glass bubble is not allowed to touch with hard substances. Electrode wills no effective if any rubbing is made.

敏感玻璃球泡不要与硬物接触,任何破损和擦毛都会使电极失效。

3. When measurement finished, put electrode in the cap, this has a bit of KCL solution in it.

完成测量后,应将电极保护帽套上,帽内应放少量氯化钾溶液。

4. Outer reference solution of combination electrode is 3 mol/L KCL solutions (Mix 20ml distilled water with 3 mol/L KCL). Solution can be poured from small hole under the cap.

复合电极的外参比补充液为3 mol/1 KCL 20ml蒸馏水和3 mol/L KCL混合),补充液可从上端小孔加入。

5. Plug of electrode must be kept clean and dry. Short circuit of two ends of the plug is prohibited.

电极的插头必须保持清洁和干燥,**防止输出端两端短路。

6. Electrode should be matched with pH meter that has high input impedance(≥1012Ω)

电极应与高输入阻抗(≥1012Ω)pH计配套。

7. Electrode mustn’t be dipped in distilled water, protein solution&nbs******bsp;acid fluoride solution for long time. It mustn’t’ touch with organic silicon grease.

电极避免长期浸在蒸馏水、蛋白质、酸性氟化物溶液中,并防止和有机硅油脂接触。

8. If slope drops, dip sensitive bubble of the electrode in 4% HF solution for 3-5 sec. clean with distilled water, dip in KCL solution to make it renew. 

如果斜率下降,把电极敏感球泡端浸泡在4%HF(氢氟酸)3-5秒,用蒸馏水洗净,浸泡在KCL溶液中使之复新。

9. If bubble is contaminated&nbs******bsp;liquid junction is blocked, sensitivity of electrode will be influenced. Clean with proper detergent.

被球泡受污染或液接界堵塞,电极灵敏度会受影响,此时应用适当清洁剂清洗。

10. Flowing table is reference for detergent selection:

清洁剂选择看下表,供参考

Contaminated Substance污染物

Detergent清洗剂

Inorganic metal oxide有机金属氧化物

Less than 1 mol/1 acid   低于/M盐酸

Organic grease substance有机脂类物质

Weak alkalinity稀皂液或洗涤剂

Resin substance树脂高分子物质

Alcohol, acetone, ether酒精,丙酮 ,乙醚

Protein blood deposit蛋白质血球沉淀物

Acidity enzyme solution酸性酶溶液

Dyestuff substance染料类物质

H2O2 diluted bleaching powder 过氧化氢,稀漂白液

Effective storage time:12 months

有效贮存时间:12个月

e-201-c-9PH复合电极英文版说明书