Withstanding Voltage Test
Output mode
AC
AC+DC
Outpu Voltage
AC:0.1~5.00KV(DC 6KV),digital setting 10V/step adjustable with UP/DOWN keys
V-setting Accuracy
± (1% of setting +5 volts) without load
Wave-form
Sine-wave
Load Regulation
1%+5volts or better
Output circuit
Linear amplifier
Voltage Limiter
Output voltager trip function
V-display
0~6.000KV F.S
V-display Accuracy
± (1% of reading +5 digits)
|-display
0.001~15mA F.S(DC 7.50mA)1μA Resolution
|-display Accuracy
GO/NG judgment
Window comparator system
limit Value
Hi:0.001~15.00mA (DC 7.50mA) Lo:Less Than Hi Limit
Judgment Accuracy
± (1% of High Limit +1μA )
Arcing value
1mA~15mA,ON/OFF(DC 6mA)
Test Time
0.1~999sec (±20mS),off
Ramp Time
Insulation Resistance (Option for 9072A)
Measuring Voltage
50~1000VDC,positive polarity (digital setting 1V/step)
± (5% of setting +5 A )
Terminal Voltage
Same the setting voltage
Measuing Range
10~9999MΩ Spen.to 2000MΩ
Measuing Accuracy
≥500V,10~2000MΩ:±10% of reading +10 digits <500V,10~2000MΩ:±15% of reading +10 digits
Limit Value Range
Both High and Low limit value can be set within the measuring range
same the measuring accuracy
Judgment time
0.3sec
0.1~999 sec (±20mS),Off
Power Requirements
Line Voltage
100V/120V/220V/240V ±10% 50 or 60Hz
Power Consumption
When no-load (Ready state):≤60W,When with Rated Load:≥300W
Insulation Resistance
≥30MΩ,with 500V DC
Withstanding Voltage
1000V AC,for 1 minute
Dimensions
270(W)×350(D)×100(H)mm
Weight
Approx. 12Kg