Performance
Measuring Range
0.5to2ppm2to12ppm12to30ppm
Number of Pump Strokes
421
Correction Factor
0.2512.5
Sampling Time2 minutes per pump stroke
Detecting Limit0.2 ppm (n=4)
Colour ChangeWhite → Pale purple
Reaction PrincipleCHCl3 + I2O5 + H2S2O7 → Cl2
Cl2 + (CH3C10H5NH2)2→ Pale purple products
Coefficient of Variation10% (for 2 to 4 ppm), 5% (for 4 to 12 ppm)
Shelf Life1 Year
Corrections for temperature & humidityTemperature correction is necessary
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.
Possible coexisting substances and their interferences
SubstanceConcentrationInterferenceChange colour by itself to
Chlorine NoNo
Bromine NoNo
Iodine NoNo
Unsaturated halogenated HCs +Pale purple
Saturated halogenated HCs +Pale purple
Calibration gas generationDiffusion tube method
TLV-TWATLV-STELExplosive range
10 ppm--
 
仕様
目盛範囲:2~12ppm〔基準n=2〕
測定範囲:0.5〔n=4〕~30ppm〔n=1〕
吸引回数〔n〕:4,2〔基準〕,1
1回(100ml)の吸引時間:2分
検知限度:0.2ppm〔n=4〕
変色:白色→淡紫色
温?湿度補正:温度
有効期限:1年(冷蔵庫保存)
指示精度:CV=5~10%
(CV:変動係数=σ:標準偏差÷平均値×100)
干渉ガス
ガス名共存の場合の濃度と結果単独の場合
塩素 影響しない変色しない
臭素 影響しない変色しない
ヨウ素 影響しない変色しない
不飽和ハロゲン化炭化水素 +誤差淡紫色に変色
飽和ハロゲン化炭化水素 +誤差淡紫色に変色
校正用ガスガス拡散管法
許容濃度管理濃度TWASTEL爆発範囲
3ppm3ppm10ppm----
标题:
内容:
联系人:
联系电话:
Email:
公司名称:
联系地址:
 
 
注:1.可以使用快捷键Alt+S或Ctrl+Enter发送信息!
2.如有必要,请您留下您的详细联系方式!
关于我们| 易展动态| 易展荣誉| 易展服务| 易家文化| 英才集结号| 社会责任| 联系我们

备案号:粤ICP备11010883号| 公安机关备案号:44040202000312| 版权问题及信息删除: 0756-2183610  QQ: 服务QQ

Copyright?2004-2017  珠海市金信桥网络科技有限公司 版权所有

行业网站百强奖牌 搜索营销*有价值奖 中小企业电子商务**服务商
以上信息由企业自行提供,信息内容的真实性、准确性和合法性由相关企业负责,易展仪表展览网对此不承担任何保证责任。
X
选择其他平台 >>
分享到