MODEL
|
5050
|
5052
|
5051
|
5053
|
AC WITHSTAND VOLTAGE
|
Output Rating
|
5KVAC/12mA
|
|
Range
|
Resolution
|
Accuracy
|
OutPut Voltage,KVAC
|
0~5.00
|
0.01
|
±(2% of setting +5V)
|
OutPut Frequency
|
50Hz/60Hz±100ppm, User Selection
|
Output Waveform
|
Sine Wave, THD<2%(Resistive Load),Crest Factor=1.3~1.5
|
Output Regulation
|
±(1% of output +5V), Form no load to full load
|
SETTINGS
|
Hi-Limit AC Current,mA
|
0.10~12.00
|
0.01
|
±(2% of setting +2counts)
|
Lo-Limit AC Current,mA
|
0.00~12.00
|
0.01
|
±(2% of setting +2counts)
|
Ramp Time, Second
|
0.1~999.9
|
0.1
|
±(0.1%+0.05sec)
|
Dwell Time, Second
|
0, 0.2~999.9(0=continuous)
|
0.1
|
±(0.1%+0.05sec)
|
Arc Detection
|
0, 1~9 ranges(0=OFF, 9 is the most sensitivity)
|
DC WITHSTAND VOLTAGE (5051,5053 Only)
|
Output Rating
|
6KVDC/5mA
|
Output Voltage, KVAC
|
0~6.00
|
0.01
|
±(2% of setting +5V)
|
Output Ripple
|
<5%(6KV/7500μA at Resistive Load)
|
SETTINGS
|
Hi-Limit DC Current, mA
|
0.02~5.00
|
0.01
|
±(2% of setting +2counts)
|
Lo-Limit DC Current, mA
|
0.00~5.00
|
0.01
|
±(2% of setting +2counts)
|
Ramp Time, second
|
0.1~999.9
|
0.1
|
±(0.1%+0.05sec)
|
Dwell Time, second
|
0, 0.2~999.9(0=continuous)
|
0.1
|
±(0.1%+0.05sec)
|
Arc Detection
|
0, 1~9 ranges(0=OFF, 9 is the most sensitivity)
|
Discharge Time
|
≤200 msec
|
INSULATION RESISTANCE(5052,5053)
|
Output Rating
|
1KVDC/1000MΩ
|
Output Voltage,KVAC
|
0.10~1.00
|
0.01
|
±(2% of setting +3V)
|
SETTING
|
Hi-Limit Resistance, MΩ
|
0.10~1000(0=OFF)
|
1
|
±(3% of setting +2counts)at ≥500VDC
|
Lo-Limit Resistance, MΩ
|
1~1000
|
1
|
±(7% of setting +2counts)at <500VDC
|
Dwell Time,second
|
0, 0.5~999.9(0=continuous)
|
0.1
|
±(0.1%+0.05sec)
|
GENERAL
|
Input Voltage AC
|
115/230 Vac±15%,50/60Hz ±5%, Fuse 3.15A Fast
|
PLC Remote Control
|
Input: Test, Reset Output: Pass, Fail, Processing
|
Memory
|
5 Sets Store/Recall of each Memory, 4 Test Mode Selectable (W,I,W~I,I~W Only 5052,5053)
|
Display
|
128X64 LCD With Back Light
|
Key Lock
|
To prevent unauthorized alteration of the test Parameters
|
Calibration
|
Build-in software and external calibrated meters
|
Environment
|
0~40℃, 20~80% RH
|
Dimension/Net Weight
|
280mm(W)×89mm(H)×370mm(D)/9kg
|
STANDARD ACCESSORIES
|
Power Cord (10A)
|
×1
|
Hipot Test Lead, 1.5m(5001)
|
×1
|
Hipot Return Lead, 1.8m(5002)
|
×1
|