MODEL
|
ESC-125
|
Voltage
|
115/230VAC±15%, User selection
|
Frequency
|
50/60Hz±5%
|
Fuse
|
2A / 250V Slow-Blow fuse
|
Line condition
|
Power Switch
|
Reverse polarity switch setting select ON / OFF
|
Neutral Switch
|
Neutral switch on/off selection for single fault condition
|
Probe Setting
|
Ground to Line (G-L)
|
V
|
Surface to Line (PH-L)
|
-
|
Surface to Surface (PH-PL)
|
-
|
Ground to Neutral (G-N)
|
V
|
Auto Function(G-L & G-N)
|
V
|
SETTINGS
|
Leakage HI/LO
Trip
|
Range
|
0uA, 1uA~10000uA (0=OFF)
|
Resolution
|
1uA
|
Leakage Current RMS Display
|
Range
|
MD Major Resistance is 0.5kΩ
|
MD Major Resistance is 1kΩ
|
MD Major Resistance is 1.5kΩ
|
Range 1
|
0.0uA - 260.0uA
|
0.0uA - 130.0uA
|
0.0uA - 85.0uA
|
Range 2
|
240.0uA - 999.9uA
1000uA - 1050uA
|
120.0uA - 525.0uA
|
80.0uA - 350.0uA
|
Range 3
|
800.0uA - 999.9uA
1000uA - 4200uA
|
400.0uA - 999.9uA
1000uA - 2100uA
|
266.6uA - 999.9uA
1000uA - 1400uA
|
Range 4
|
3600uA - 10000uA
|
1800uA - 10000uA
|
1200uA - 10000uA
|
Leakage Current RMS Display
|
Range
|
Frequency
|
Basic Accuracy
|
Range 1-4
|
DC, 15Hz ~ 200kHz
|
±(2% of reading + 5 counts)
|
200kHz ~ 1MHz
|
±(5% of reading)
|
MD Circuit module
|
Measuring Device(MD)
|
A, E, F, G
|
MD A. UL544 Non Patient, UL484
|
MD E. UL60950, IEC60950, IEC61010-1, IEC60335-1, IEC60990
|
MD F. External MD
|
MD G. Basic measuring element 1KΩ of Frequency check
|
MD Components
|
Resistace accuracy = 1%; Capacitance accuracy = 5%
|
MD Voltage Limit
|
Maximum 32Vdc
|
Leakage Current Offset
|
Range: 0-999.9uA; Resolution: 0.1uA
|
DUT Power
|
AC Voltage Display
|
Range
|
30.0 - 300.0V
|
Resolution
|
0.1V
|
Accuracy
|
±(1% of reading + 2 counts)
|
AC Current
|
30A max continuous
|
Over current protection
|
32A, Response time < 600ms / 68Apeak, Response time < 600ms
|
Delay Time
|
Range
|
0, 1.0 - 999.9s (0=continuous)
|
Resolution
|
0.1s
|
GENERAL
|
PLC Remote Control
|
Input : Test, Reset, and Recall Memory 1 through 3
|
Output:- Pass, Fail, Processing, Start-Out, Reset-Out
|
Memory
|
99 memories, 8 setps / memory
|
Display
|
20 x 2 OLED
|
Alarm Volume Setting
|
Ranger: 0-9; 0=OFF, 1 is softest volumne, 9 is loudest volume
|
Key Lock
|
To prevent unauthorized alteration of the test Parameters
|
Calibration
|
Software and adjustments are made through front panel
|
Interface
|
optional USB & RS232, GPIB
|
Environment
|
0-40℃, 20-80% RH
|
Dimensions(W×H×D), mm
|
430×89×400
|
Net Weight
|
11.5kg
|
*product specifications are subject to change without notice.
|