AC WITHSTAND VOLTAGE |
|
Range |
Resolution |
Accuracy |
Output Voltage, Vac |
0 - 5000 @ 40mA max |
1 |
±( 2% of setting + 5V) |
Output Frequency |
50 or 60 Hz, ?00 ppm |
Output Waveform |
Sine wave, crest factor > 1.3 and < 1.5 |
Output Regulation |
±1 % of setting + 5V, from no load to full load |
SETTINGS |
|
|
|
Hi Limit, current, mA |
0.00 - 40.00 |
0.01 |
±( 2% of setting + 2 counts) |
Lo Limit, current, mA |
0.000 - 9.999 |
0.001 |
±( 2% of setting + 2 counts) |
Dwell Time, second |
0.3 - 999.9 |
0.1 |
±( 0.1% + 0.05 sec.) |
Ramp Time, second |
0.1 - 999.9 |
0.1 |
±( 0.1% + 0.05 sec.) |
Offset, current, mA |
0 - 2.000 or Auto |
Acr Detecion |
9 ranges |
DC WITHSTAND VOLTAGE |
Output Voltage, DCV |
0 - 6000 |
1 |
±( 2% of setting + 5V) |
SETTINGS |
|
|
|
Hi-Limit, current, μA |
0 - 9999 |
1 |
±( 2% of setting + 2 counts) |
Lo-Limit, current, μA |
0 - 999.9 |
0.1 |
±( 2% of setting + 2 counts) |
Dwell Time, second |
0.4 - 999.9 |
0.1 |
±( 0.1% + 0.05 sec.) |
Ramp Time, second |
0.1 - 999.9 |
0.1 |
±( 0.1% + 0.05 sec.) |
Offset, current, μA |
0 - 200.0or Auto |
Ramp-High, current, mA |
On/Off user selection (12mA peak max.) |
Charge-Low, current, μA |
0 - 350 or Auto |
INSULATION RESISTANCE |
|
Range |
Resolution |
Accuracy |
Output Voltage, Vdc |
100 - 1000 |
1 |
±( 2% of seffing+2V) |
SETTINGS |
Hi Limit, resistance, M Ω |
1 - 9999, 0=off |
1 |
±( 2% of setting + 2counts) @500-1000VDC/1-1000 M Ω |
Lo Limit, resistance, M Ω |
1 - 9999 |
1 |
±( 8% of setting + 2counts) @100-1000VDC/1-9999 M Ω |
Delay Time, second |
0.5 - 999.9 |
0.1 |
±( 0.1% + 0.05 sec.) |
Charge-Low, current, μA |
0 - 3.500, or Auto |
GROUND BOND |
Output Current, AC |
3.00 - 30.00 |
0.01 |
± (2% of reading + 2 counts) |
Output Voltage, AC |
3.00 - 8.00 |
0.01 |
±( 2% of reading + 3 counts) |
Output Frequency, Hz |
50 or 60, ± 100 ppm |
SETTINGS |
|
|
|
Lead Resistance Offset, m Ω |
200 |
1 |
±( 1% of setting + 2 m |
Hi Limit, resistance, m Ω |
0 - 600 |
1 |
±( 2% of setting + 2 m |
Dwell Time, second |
0.5 - 999.9 |
0.1 |
±( 0.1% + 0.05 sec.) |
MEASUREMENT |
Voltage, KV |
0.00 - 6.00 |
0.01 |
±( 2% of reading + 1 count) |
AC Current, mA |
0.000 - 3.500 3.00 - 40.00 |
0.001 0.01 |
±( 2% of reading + 3 counts) |
DC Current, μA |
0 - 350.0 300 - 3500 3000 - 9990 |
0.1 1 10 |
±( 2% of reading + 3 counts) |
DC Voltage, V(Insulation) |
0 - 1000 |
1 |
? ±( 2% of reading + 2 counts) |
Resistance, M Ω |
1 - 9999 Autorange |
- |
±( 2% of setting + 2counts) @500-1000VDC/1-1000 M Ω |
±( 8% of setting + 2counts) @100-1000VDC/1-9999 M Ω |
GENERAL |
Memory |
50 sets, 8 steps per memory |
PLC Remote Control |
Test, Reset, Recall memory 1, 2, 3 functions |
PLC Signal Output |
Pass, Fail, Processing |
External Scanner Port |
Two, for connecting to 7002 scanners |
Internal Scanner |
Optional, 8 channels, factory installed |
LCD Display |
20 x 2 LCD with back light, 9 ranges contrast setting |
Key Lock |
Front panel or 4-digit password |
Pass/Fail Indicator |
Lighted button and settable audible alarm |
Calibration |
Build-in software and external calibrated meters |
Interface |
RS232 standard, GPIB and Printer port optional |
Dimension / Net Weight |
EIA 3U /430(W) x 133(H) x 500(D) mm / 22 Kg. |
Input Voltage, AC |
115/230 V ± 15%, 47 - 63 Hz |
Environment |
0 - 40 ℃ , RH 0 - 85% |