Performance Hydrogen Sensitivity Range: 0.4% to 5% hydrogen by volume at 1 ATM Typical Response Time: T90 of 60 sec. maximum Ingress Protection: IP64 capable Recommended Verification Interval: 90 days Product Life Expectancy: 10 years Accuracy: ± (0.03 x indication + 0.2) percent Hydrogen-by-volume
Interface Options Power Barrier Input Voltage: 24 VDC nominal Power Barrier Input Voltage Range: 20 VDC to 28 VDC Intrinsically Safe Sensor Input Voltage: 10 VDC Sensor Input Voltage Range: 5 VDC to 28 VDC Input Power: 10 W Analog Output: 4 mA to 20 mA Serial Communication Options: RS422
Operating Conditions Process Gas Temperature Range: -20ºC to 65ºC Process Gas Humidity: 0% to 100% Gas Flow Rate: 0.1 slpm to 50 slpm Gas Pressure: 0 to 7 bar gauge, 0 to 100 psi gauge Operating Temperature Range: -20ºC to 40ºC Operating Humidity: 0% to 95% RH (non-condensing) Storage Temperature Range: -40ºC to 50ºC Calibration Background Gas: Air
Included Accessories Intrinsically Safe Power Barrier Armored Power / Analog Cable: 4 m Calibration Cup
Optional Accessories Single Channel Intrinsically Safe Analog Output Barrier (with or without relays) Single Channel Intrinsically Safe RS422 Serial Interface Barrier Serial Interface Cable: 4 m (custom lengths available) Armored Power / Analog Cable (custom lengths available)
Adapter Fittings ½ in. MNPT thread ½ in. FNPT thread -8 SAE/MS thread size ... and many other industry standards.
Dimensions Total Length: 10.3 in. Length of Enclosure: 6.8 in. Length of Tube: 2.9 in. Width: 3.4 in. Depth: 1.4 in. H-Tech Shipping Information Estimated Ship Dimensions: " x " x " • Estimated Ship Weight: lbs • Lead time: 14 days • Shipping origin: Valencia, United States Please contact us for pricing on this product. H-Tech's Shipping Policy