可测量6个参数:pH, ORP, EC, TDS,NaCl百分比,多种温度量程测试 PH可在7个记忆点人选三个自动校准,甚至在测试两种量程差别很大的试样时也能提供的校准曲线。在EC 和 TDS测量时具有自动量程修正特点,可为测试样品自动调节分辨率。测量可以设定在20度和25度,使用者可选择自动温度补偿。在测量电导率时可以关闭自动温度补偿。屏幕上的稳定性指示器可使测量*加。为测量电导率提供4环电导电。GLP特点可以记录,储存数据并通过RE232或USB传到电脑中
参数
MTYK-MI180
范围 PH
MV
EC
TDS
NACL
温度
-2.00 to 16.00 pH; -2.000 to 16.000 pH
±699.9 mV; ±2000 mV
0.00 to 29.99 μS/cm; 30.0 to 299.9 μS/cm; 300 to 2999 μS/cm;
3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm (uncompensed EC*)
0.0 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm); 150 to 1499 mg/L (ppm);
1.50 to 14.99 g/L (ppt); 15.0 to 100.0 g/L (ppt);
up to 400.0 g/L actual TDS (with 0.80 factor)
0.0 to 400.0%
-20.0 to 120.0°C / -4.0 to 248.0°F
分辨率 PH
0.01 pH; 0.001 pH
0.1 mV; 1 mV
0.01 μS/cm; 0.1 μS/cm; 1 μS/cm; 0.01 mS/cm; 0.1 mS/cm;
0.01 mg/L; 0.1 mg/L; 1.0 mg/L; 0.01 g/L; 0.1 g/L
0.1%
0.1°C / 0.1°F
精度 PH
±0.01 pH; ±0.002 pH
±0.2 mV; ±1 mV
±1% of reading ±(0.05 μS/cm or 1 digit)
±1% of reading ±(0.03 ppm or 1 digit)
±1% reading
±0.4°C / ±0.8°F
REL MV OFFSET
±2000 mV
校准 PH
1, 2 or 3 点校准, with 7 memorized buffers (pH 1.68, 4.01, 6.86, 7.01, 9.18,
10.01 and 12.45)
1 point slope calibration with 6 memorized solutions: (84 μS/cm, 1413 μS/cm,
5.00 mS/cm, 12.88 μS/cm, 80.0 μS/cm, 111.8 mS/cm)
1 point, with MA9066 solution
2 point, at 0 and 50°C / 32 and 122°F
温度补偿
自动或手动 -20.0 to 120.0°C / -4.0 to 248.0°F
温度系数
0.00 to 6.00%/°C (EC and TDS only)进行选择
PH及温度电
MA917B/1 MA831R
EC/TDS/NACL/温度电
MA814D/1
TDS因子
0.40 to 0.80 (default value is 0.50)
LOG ON DEMAND
up to 50 samples on each range (pH, mV, EC, TDS, NaCl)
GLP
上一个pH, EC, NaCl校准数值
PC转换接口
RS232或USB光学转换
环境
0 to 50°C / 32 and 122°F; max RH 95%
输入阻抗
1012 Ohm
电源
12 VDC power adapter (included)
尺寸
230 x 160 x 95 mm
重量
0.9 kg