Yokogawa Conductivity Sensor SC210G-A-100-L015-05*A
General Specifications:
SC210G:
Cable with ring terminals (applicable to FLXA21, SC202G/TB and SC202SJ/TB)
Cable with M3 ring terminals (applicable to SC450G, SC202口/TB)
Cable with pin terminals (applicable to FLXA21,SC202G and SC202SJ)
Cable with M4 ring terminals (applicable to FLXA21)
Object of measurement: Conductivity of solutions
Measuring principle: Two-electrode system
Yokogawa Conductivity Sensor SC210G-A-100-L015-05*A
Cell constant : 0.05 cm-1, 5 cm-1
Measuring range: 0 - 0.5 μS/cm to 0 - 200 μS/cm (Cell constant: 0.05 cm-1)
0 - 200 μS/cm to 0 - 20 mS/cm (Cell constant: 5 cm-1)
Temperature Range: 0 to 105 C (chamber material: SCS14)
0 to 100 C (chamber material: Polypropylene)
Pressure range: 0 to 1 MPa (chamber material: SCS14)
0 to 500 kPa (chamber material:Polypropylene)
Measuring solution condition for Yokogawa Conductivity Sensor SC210G-A-100-L015-05*A
Although flow rate is not limited in measurement, less than 20 l/min is recommended for flow-through type. If slurry is included in sample solutions in flow-through type detectors, the electrode part and the inside of solution chamber may be worn significantly.
Air bubbles should not be mixed in the sample solutions to obtain correct measured values.
Temperature sensor: Thermistor (PB36NTC).
Wet part Materials for Yokogawa Conductivity Sensor SC210G-A-100-L015-05*A
SC210G: For sensor, SUS 316 stainless steel, Viton (O-ring) and Polytrifluorochloroethylene
For body, SUS316 stainless steel, polypropylene and Viton (O-ring)
SC210G :For sensor, Platinum, glass and Viton (O-ring)
For body, SUS316 stainless steel, polypropylene and Viton (O-ring)
Construction:JIS C0920 watertight (equal to NEMA 4)