产品详情
  • 产品名称:Hemolysis test kit

  • 产品型号:#: K003
  • 产品厂商:
  • 产品价格:0
  • 折扣价格:0
  • 产品文档:
你添加了1件商品 查看购物车
简单介绍:
品名 货号 规格 Hemolysis test kit #: K003 Hemolysis test kit (30 determinations)
详情介绍:

品名Hemolysis test kit

货号#: K003

规格Hemolysis test kit (30 determinations)

简介

"PRODUCT #: K003


Hemolysis assay

Hemolytic activity is a requirement to be tested for any blood contacting medical device


Hemolysis test kit (30 determinations)

European Union Price: €395,- (As per 01-01-2018).

Inquire for a quotation if you are located outside the European Union.



Order test kit Hemolysis test kit

Description

Hemolytic activity is a requirement to be tested for any blood contacting medical device. The test is based on erythrocyte lysis induced by contact, leachables, toxins, metal ions, surface charge or any other cause of erythrocyte lysis. The current description is based on direct contact of biomaterial and an erythrocyte suspension. The method is based on release of hemoglobin, which can be measured spectrophotometrically. This method is suited to evaluate the hemocompatibility of biomaterials and medical devices according to the international standard ISO 10993-4:2002.


Application

Measuring biomaterial induced hemolysis. This kit is intended for laboratory research use only and is not for use in diagnostic or therapeutic procedures. The analysis should be performed by trained laboratory professionals.


Principle

An erythrocyte suspension is incubated for 24 hours with test material during rotation at 37 ºC. Before and after incubation samples are collected and centrifuged to obtain supernatant, containing free hemoglobin. The hemoglobin concentration is measured by means of a spectrophotometer. Test samples are compared to reference materials. Positive reactive and less-reactive reference materials are included in the kit. It is recommended to include at least two reference materials in each analysis.

The results of the tested materials in relation to the reference materials may be used to evaluate the hemolytic activity. It must be noted that pass/ fail criteria are based on 2% hemolysis. Thus, also the total hemoglobin concentration of the used erythrocyte suspension must be determined.

The kit is designed to determine hemolytic activity of small biomaterial samples. Larger samples can also be used as long as the ratio between erythrocyte suspension and material size is respected.


Storage and Stability

Product is stable at -20 °C for at least six months.


Download

Manual / SDS


References

Nemani KV, Moodie KL, Brennick JB, Su A, Gimi B. In vitro and in vivo evaluation of SU-8 biocompatibility. Mater Sci Eng C Mater Biol Appl. 2013 Oct;33(7):4453-9."



标题:
内容:
联系人:
联系电话:
Email:
公司名称:
联系地址:
 
 
注:1.可以使用快捷键Alt+S或Ctrl+Enter发送信息!
2.如有必要,请您留下您的详细联系方式!
相关文章

沪公网安备 31011202007343号