Specification
|
DSD-4M
|
Measuring range
(at the time of Track)
|
0.030 - 4 N-m
|
0.30~40 kgf-cm
|
0.30~35 lbf-in
|
*A measurement unit is fixed to one unit. Specify units when ordering
|
Accuracy
|
± 0.5% (in case of reading 499 or less = last digit +/- 3 reading)
|
Measuring unit
|
lbf・in-kgf・cm-N・m
|
Display
|
The 4 figures digital display of LCD
|
The measurement direction
|
CW-CCW (right and left)
|
Measuring
mode
|
P-P Peak
|
Holding Peak Load Value
|
T-R Track
|
Showing Real time torque value ( Used for Zero-Adjustment )
|
P-D Peak down
|
Holding Load Value at the moment to change from upward to downword
|
C Real-time output
|
Load torque value is outputted every about1/ 70 second by RS232C.
|
Decision
function
|
Setting of torque value
|
Programmable High and Low setpoints with both audible beep and Green/Red LED indicator for uniform torque tightening or GO/NO GO testing
|
The Maximum, the minimum, and the average value
|
Displays, memory and output the data number, the maximum, the minimum, and the average value of the measured value to 30 data. It can save by a maximum of 10 channels.
|
Data spec.
|
Data transfer
|
Serial interface RS232C (Baud rate : 9600)
|
Data memory
|
300 data / ten channels data
|
A setup of the number of counts
|
One to 99 count (count in a direction with a bundle)
|
Power supply
|
Ni-Cd chargeable battery 1.2V×4cells 600mAh
|
Auto power save
|
10 minutes no use, automatically switch off.
|
Charging time
|
From empty - less than 5hours
|
Continues working time
|
12 hours
|
Bit fit mouth
|
HEX6.35
|
Weight
|
330g
|
Attachment
|
Bit (+) #1 #2
|
Accessories
(one piece each)
|
Dedicated AC/DC Adaptor (Charger), 230VAC or 120VAC- DC 7.5V 120mA Output
|
An inspection report, a proofreading certificate, a traceability system figure
|